This blog post will cover the basic principles behind eye
tracking, the customization done to our Avotec eye tracking system, and it’s use
in conjunction with Arrington’s Viewpoint software to obtain simultaneous eye
tracking data during an fMRI experiment.
Why collect eye
tracking data?
The use of simultaneous eye tracking is common in vision
fMRI experiments where frequent eye movements are expected, but it can also be
a useful way to monitor alertness in your subject. Resting-state fMRI scans in
particular, where it is generally assumed that the subject is awake with their
eyes open, can benefit greatly from monitoring subject eye movement in real-time.
Catching your subject asleep two minutes into a ten-minute EPI can be disheartening,
but it’s better than finding out after the fact that you wasted 10 minutes and their
data is unusable. (See Note 1)
How does eye tracking
work?
In a typical system, near infrared (IR) light is used to
produce corneal reflections that can be monitored and recorded by a camera,
producing a time series of the pupil diameter for the target eye (See Note 2). For some applications, such as
compliance with instructions during a resting-state scan, sufficient
information may be obtained from a simple bore-mounted optical camera directed
at the subject’s face. In either case, however, it is necessary to ensure that
the monitoring system does not interfere with visual stimuli or other
peripheral equipment essential for fMRI, a requirement that can be challenging
in the cramped space in and around a typical MR head coil. The Avotec unit is
not affected by these limitations because of its design, though we still
modified the unit in order to optimize performance. Our EyeLink unit (which may
be the subject of a future post) however did have to be heavily modified before
it could even be used in our facility.
The Avotec RE-5601
Eye Monitoring System
Avotec’s system uses special mirrors that allow IR light to
pass through the front of the carriage whilst simultaneously allowing the
subject to view images on a rear-facing projector system. Their RE-5601 model
has since been discontinued and replaced with the RE-5700 unit,
so the modifications we’ve done may not be required/ideal/possible with that
unit.
Figure 1 - Original Avotec mirror system (left) and the mirror system after our modifications (right) |
The most obvious modification made was completely scrapping
the carriage and re-making it in black nylon; which is a bit sturdier and the opacity
eliminates ambient light interference. The green felt on the sides minimize
unwanted reflections and the lens of the camera was replaced with an after
market one from Avotec. Design changes were made to the separate camera stage
so as to allow for a greater degree of freedom to adjust the camera and IR source
independently of each other. The basic idea of Avotec’s design remains intact
but the changes made led to a significant improvement in image quality and a
reduction in setup time.
Figure 2 - Modified Avotec mirror and camera on 12 Ch. Siemens coil |
Figure 2 features the mirror system and camera attached
to our 12 Ch. coil. The camera and IR source cables are wrapped in a protective
plastic sheath to help minimize strain on the sensitive fiber optic cables. Although
not common, it is also possible to use this mirror for non-eye tracking
experiments by leaving the camera unit off. Figure 3 shows our custom mirror for the 32 Ch coil.
It’s the same basic idea, with a base to secure the camera and IR sources, but
several other factors had to be considered when designing it to fit within the
bore of the magnet.
Figure 3 - Modified Avotec mirror for 32 Ch. Siemens coil |
The eye tracker interface unit sits on the side of the bore
during experiments, see Figure 4. Nylon casings were fixed onto the
connectors and an extended platform (where the camera is sitting in the photo) were
added after several incidences of the unit being damaged and cables coming
loose. This might seem like overkill, but sometimes it’s necessary for a piece
of equipment that contains fiber optics and gets used as often as this.
Figure 4 - Eye tracking camera interface with support structure and modifications |
Setting up the Avotec
camera and recording
Even before we made any of our modifications, the Avotec
system was quick to set up. The mirror slides easily into position on the coil
and it is just a matter of finding the subject’s eye, focusing on it and
adjusting the IR illumination to eliminate as much shadow as possible. The issue of shadows on the subject’s face can be a problematic
source of frustration, however. Shadows cause the camera to lose sight of the
subject’s eye, figure below, so adjusting to eliminate as much as possible from
the get go is strongly recommended even though it can take some time and fine-tuning. Our
modifications allow some adjustment to the camera and IR source,
so I recommend allocating at least 10 minutes to eye tracking setup and
optimization. The position and lighting
shown in the figure below is an accurate representation of the image quality
required for obtaining proper eye tracking data.
Figure 5 - Example of excessive (left) and acceptable (right) shadow |
A video demonstration of both acceptable and unacceptable shadow can be seen by clicking on the corresponding links (see Note 3).
The Avotec unit came with Arrington’s Viewpoint software. Once the subject’s eye placement and illumination is within acceptable parameters (see Note 4), recording data is a simple matter of clicking a button at the start and at the end of a scan. A calibration sequence should be performed often (see Note 5) and it’s recommended that each scan be recorded separately to keep the file sizes manageable. Figure 6 is a screenshot of the program and the sort of information that is displayed and collected.
The Avotec unit came with Arrington’s Viewpoint software. Once the subject’s eye placement and illumination is within acceptable parameters (see Note 4), recording data is a simple matter of clicking a button at the start and at the end of a scan. A calibration sequence should be performed often (see Note 5) and it’s recommended that each scan be recorded separately to keep the file sizes manageable. Figure 6 is a screenshot of the program and the sort of information that is displayed and collected.
Figure 6 - Viewpoint software screenshot |
Don’t forget Voltage
Pro!
The final piece of the puzzle is easily one of
the most important. Arrington also sells a separate add-on, though it should
really come standard with the system, package called Voltage Pro that is
absolutely essential for fMRI and eye tracking data.
Figure 7 - Voltage Pro Add-on |
Once installed, Voltage Pro inserts a marker into the eye
tracking data set every time it receives a TTL signal from the MRI equipment.
This allows researchers to sync up eye tracking data to the start of data acquisition
with relative ease, so long as the add-on is opened before recording starts. Once data is collected it is saved as a .txt
file, see below, that can easily be imported to other programs for analysis.
And that’s it! Hopefully this encourages more users to
consider acquiring such useful data. While significant modifications were done
to the unit, it is still quite recognizable as the same system. The same cannot
be said, however, for the Eyelink unit that we purchased some years back. But
that is a post for another day!
Note 1: An even
worse scenario would be unknowingly using the sleepy subject data set in your
study!
Note 2: Some eye
trackers are capable of capturing data from both eyes. Our Avotec system does
not have this option.
Note 3: These videos were intentionally taken before any calibration was done, since the primary focus was to showcase the issue of shadow, and I was unfortunately not able to zoom in on the image of the eye before posting. Because of this all information present on the screen other than the video image should be disregarded.
Note 4: From
previous experience, research groups have their own parameters for what is and
is not considered acceptable when acquiring eye tracking data.
Note 5: my
recommendation is to run it before every scan, but time does not always allow
for that.